Link:
Pharm/Biotech Resources Released by FDA: 8/31/06. Posted by FDA: 9/11/06 Stephen P. DeFalco
Between March 6 and 24, 2006, Barbara J. Breithaupt, Sriram Subramaniam, Ph.D., Martin K. Yau, Ph.D., Michael F. Skelly, Ph.D., Nilufer M. Tampal, Ph.D., John A. Kadavil, Ph.D., and Jacqueline A. O'Shaughnessy, Ph.D., representing the Food and Drug Administration (FDA), conducted a follow up inspection of several bioequivalence studies performed by MDS Pharma Services (MDS) in Saint Laurent (Montreal), Quebec Canada, including the following:
These inspections are a part of FDA's Bioresearch Monitoring Program, which includes inspections designed to evaluate the conduct of research, to confirm that data intended for FDA submission is reliable for FDA regulatory decisions, and to verify compliance with Title 21 of the Code of Federal Regulations (CFR), Part 320, Bioavailability and Bioequivalence Requirements. Previous FDA inspections of the MDS analytical facility in Saint Laurent found significant deficiencies that raised concerns about the validity of bioequivalence data generated by MDS. Specifically, on April 26, 2004, FDA issued to Gilbert Godin, Group Vice-President, Early Stage Development, correspondence citing MDS' failure to conduct a systematic and thorough evaluation to identify and correct sources of contamination and implement adequate policies and procedures to address such contamination issues. The letter discussed FDA inspectional findings related to [redacted] Study [redacted] On December 21, 2004, FDA issued to Gilbert Godin, President and CEO, a second letter citing MDS' systemic failure to analyze and investigate anomalous testing results across multiple studies for multiple sponsors. The letter discussed FDA inspectional findings related to five [redacted] studies for three sponsors and four [redacted] studies sponsored by [redacted] Because the identified deficiencies indicated a widespread problem in your analytical laboratory, FDA recommended that MDS review the validity of bioequivalence studies you conducted within the last five years. In response to FDA's letter, your senior management team met with FDA in February 2005 and NMS agreed to conduct a retrospective review of all bioequivalence studies conducted at the St. Laurent facility for the last five years (January 2000 through December 2004) and complete the review within one year. The original MDS plan to conduct the retrospective review was approved by MDS management in March 2005. At the conclusion of the current inspections in Saint Laurent and Blainville, our personnel presented and discussed with Michael J. Butler, Ph.D. and Charles Grandmaison, respectively, the items listed on Form FDA 483, Inspectional Observations. The results of these inspections and our review of related documents lead us to conclude that you failed to demonstrate that your five year retrospective review is effective and capable of discriminating between valid and invalid study data and assure that the analytical methods used for in vivo bioavailability studies conducted in your facilities in Saint Laurent and Blainville could accurately measure the actual concentration of the active drug ingredient, or its active metabolite, achieved in the body, as required by 21 CFR 320.29(a). The details of these findings are listed below. Five Year Retrospective Review (.MDS Saint Laurent) Our inspection found numerous significant deficiencies in your retrospective review, including the following:
In addition to your failure to include Study [redacted] in the
retrospective review, our [redacted] Study [redacted] (MDS Saint Laurent) Our inspection found that the pharmacokinetic profiles for some subjects had unexpected concentration results. For example, maximum concentrations (Cmax) of [redacted] were followed by a sample with no measurable drug concentration or occurred at the first or last post-dose blood sampling time points. In addition to aberrant results at or near Cmax, there were anomalous concentrations at other time points in the pharmacokinetic (PK) profile. You failed to investigate the cause of these anomalous results, or reassay the affected samples. In your response dated June 9, 2006, your retrospective review of this study concluded that "considering the low frequency of occurrence across the entire study sample set as well as within the individual PK profile, these incongruent concentrations do not impact the overall accuracy and/or validity of the reported data." Contrary to your response, the frequency of occurrence is not a justification for accepting anomalous study results. Because of these unexplained anomalous results, we are concerned about the accuracy of the reported PK parameters (Cmax and AUC) for individual subjects based on your concentration results. Your failure to investigate anomalous results in this study is similar to previous FDA inspectional findings for numerous studies regarding [redacted] and [redacted] as discussed in the FDA letters to MDS dated April 26, 2004 and December 21, 2004. [redacted] Study [redacted] (MDS Saint Lauren) Our inspection found that the Period 1 samples from Subject 19 had internal standard (IS) responses that were 5 to 6 times the average IS response of calibrators and quality controls. You did not adequately investigate the anomalous results or reassay the affected samples. Your retrospective review dated May 19, 2006 concluded that the abnormal IS response is "subject specific" and "not due to an analytical reason," providing no documentation to support your position. Contrary to your response, the IS response for Subject 19 in Period l fails to demonstrate subject specificity because the Period 2 samples for Subject 19 did not exhibit a similar abnormally high IS response. Also, since you failed to demonstrate that a similar aberrant response occurred upon reanalysis, you lack data to support your conclusions. [redacted] Study [redacted] (MDS Saint Laurent) Our inspection found that you failed to identify the biased exclusion of individual calibration points from the standard curve in run 13. Recalculation of the standard curve in an unbiased manner during the inspection found that the run should have been rejected because the quality control (QC) samples did not meet the run acceptance criteria. Your retrospective review dated January 4, 2006 failed to identify the biased exclusion and failed to determine that the data from the run was not valid due to QC failure. [redacted] Studies [redacted] and [redacted] (MDS Blainville) FDA also inspected numerous studies fort - conducted at your Blainville
facility. Our These study-specific findings from FDA's March 2006 inspection are in addition to previous deficiencies found during FDA inspections in July 2003, February 2004, and September 2004. The previous inspections found that your analytical methods were not demonstrated to be accurate when utilized in the following ten bioequivalence studies:
In summary, the significant deficiencies regarding your five year retrospective review and failure to demonstrate the accuracy of your analytical methods in more than thirty studies for six different drugs confirm that there are widespread problems at your facilities in Saint Laurent and Blainville. Based on FDA's multiple inspections of these facilities (July 2003, February 2004, September 2004, March 2006) and our evaluation of numerous studies, we conclude that you failed to systematically investigate contamination and anomalous results, conduct an effective retrospective review, and demonstrate that your retrospective review is capable of discriminating between valid and invalid study data. If you have questions or concerns about the issues raised in this letter, please reply to:
|
||||||||||||||
|