Pharm/Biotech Resources Released by FDA: 6/25/07. Posted by FDA: 7/6/07 Paul E. Kirby, Ph.D. Dear Dr. Kirby:
This inspection is a part of FDA's Bioresearch Monitoring Program, which includes inspections designed to verify compliance with Title 21 of the Code of Federal Regulations (CFR), Part 58--Good Laboratory Practice (GLP) regulations. The regulation at 21 CFR 58 applies to nonclinical laboratory studies of products regulated by FDA. At the conclusion of the inspection, our investigator presented and discussed with you the items listed on Form FDA 483, Inspectional Observations. Following our review of the establishment inspection report and related documents, including your response dated February 19, 2007, we conclude that you violated FDA regulations governing the conduct of nonclinical laboratory studies. This letter provides you with written notice of the violations. The applicable provisions of the CFR are cited for each violation. 1. Failure of the study director to assure that all
experimental data were accurately recorded and verified [21 CFR 58.33(b)]
and document the reason for any change in the entries 121 CFR 58.130(e)]. Furthermore, dosing records indicate that animals in the two positive control groups received double the protocol-specified dose, based on the volumes documented at the time of dosing. Approximately seven weeks after dosing, the study director altered the source records to reduce the dosing volumes to one-half the volume originally documented. The study director failed to document the reason for the changes. Consequently, the actual dose administered to the animals in both positive control groups is questionable. Your study director's conclusion in the final report that the criteria for a valid assay were met based on the response from the positive controls is invalid because the dose that elicited the response cannot be assured. Contrary to your response dated February 19, 2007, while the dose volume corrections for the test and vehicle control groups were related to rounding up the decimal values of the calculated dose volumes, the corrections for the. positive control groups were not. 2. Failure of QAU to fulfill its responsibilities 121 CFR 58.35(b)]. a. Failure of QAU to assure that the reported results in the final study report accurately reflected the raw data [2l CFR 58.35(b)(6)]. For Study [redacted] the net nuclear grain count per cell was the critical measurement used to conclude the study outcome. However, we found discrepancies in the data reported for this measurement. For example,
In light of these discrepancies, we have no assurance of the quality and integrity of the final report you prepared for Study [redacted]. As described in your response dated February 19, 2007, we acknowledge your proposal to correct these discrepancies and submit a revised final report to the sponsor. b. Failure of QAU to maintain a copy of the master schedule of all nonclinical laboratory studies indexed by nature of the study [21 CFR 58.35(b)(1)]. Specifically, the master schedule did not list terminated GLP study [redacted] Additionally, you failed to maintain a key to decode the nature of studies with code c. Failure of QAU to maintain written and properly signed records of each periodic inspection [21 CFR 58.35(b)(3)]. For Study [redacted] your QAU documented inspection of colony counting on October 26, 2005. However, according to your source records, no colony counting occurred on that day; instead, the counting actually occurred on October 24, 2005. In your response dated February -19, 2007, you attributed the observations under item 2 above, to inadvertent errors on the part of the QAU. As described in your response dated February 19, 2007, we acknowledge your proposal to revise your QAU procedures for future studies. 3. Failure to identify the test and control articles-with appropriate characteristics in the final report [21 CFR 58.185(a)(4)] The final reports prepared by your study directors for Studies [redacted] did not include required information on test and control article characterization. For example, the purity and specific lot numbers of the test article and the vehicle control for Study land the purity and stability of the test article for Study [redacted] were not included. The statement in the final report that purity and composition of the test article were determined by the sponsor is not adequate to satisfy the requirement that the final report contain the actual characteristics of the test article. 4. Failure to adequately test, calibrate, and /or standardize all equipment used for the generation, measurement, or assessment of data [21CFR 58.63(a)]. You failed to assure that the [redacted] used to count the
colonies in Studies [redacted] was adequately calibrated to assure
an accurate measure of cell and/or colony counts. Although your SOP [redacted]
required instrument calibration prior to use, your records indicate that
you used this instrument without calibrating it for 15 of 20 days for
Study [redacted] and 1 of 3 days for Study [redacted] 5. Failure to prepare a final report for each nonclinical laboratory study 121CFR 58.185 (a)). Your study director failed to prepare final reports for Studies [redacted] terminated after administration of the first dose of test article. Also, we found at least six GLP studies [redacted] with draft reports pending since May 2004. The failure to prepare final reports for each nonclinical laboratory
study is a repeat violation. Following the inspection of your
facility in April 2003, FDA's letter to you dated August 25, 2003
expressed our concerns about this deficiency. In your December, 3,
2003 response you proposed to correct this deficiency by preparing
memoranda for terminated studies that would include an overview of the
study history. The current inspection found that the memorandums prepared
for terminated studies This letter is not intended to be an all-inclusive list of deficiencies at your facility. Your violation of the FDA regulations outlined above resulted in the submission of unreliable data to the sponsor. You must address these deficiencies and establish procedures to ensure that any on-going or future studies be conducted in compliance with FDA . regulations. Within fifteen (15) working days of receipt of this letter, you must notify this office in writing of the specific corrective actions you will take to address all of the deficiencies noted above and to achieve compliance with the FDA regulations. If corrective actions cannot be completed within 15 working days, you may request an extension of time in which to respond by stating the reason for the delay and the time within which the corrections will be completed. We will review your response and determine whether it is adequate. Failure to provide adequate assurances of compliance with FDA regulations may result in further regulatory action without further notice. Your reply should be sent to:
|
||||||
|